Radiol med DOI 10.1007/s11547-009-0462-y HEAD AND NECK RADIOLOGY RADIOLOGIA DEL CAPO E COLLO Radiation dose saving through the use of cone-beam CT in hearingimpaired patients Riduzione di dose radiante mediante utilizzo della TC a fascio conico nei pazienti ipoacusici N. Faccioli1 • M. Barillari1 • S. Guariglia2 • E. Zivelonghi2 • A. Rizzotti2 • R. Cerini1 R. Pozzi Mucelli1 1 Istituto di Radiologia, Policlinico G.B. Rossi, Università di Verona, P.le L.A. Scuro, 37134 Verona, Italy U.O. Fisica Sanitaria, Azienda Ospedaliera di Verona, Verona, Italy Correspondence to: N. Faccioli, Tel.: +39-045-8124301, Fax:+39-045-8027490, e-mail: [email protected] 2 Received: 22 February 2008 / Accepted: 8 July 2008 © Springer-Verlag 2009 Abstract Purpose. Bionic ear implants provide a solution for deafness. Patients treated with these hearing devices are often children who require close follow-up with frequent functional and radiological examinations; in particular, multislice computed tomography (MSCT). Dental volumetric cone-beam CT (CBCT) has been reported as a reliable technique for acquiring images of the temporal bone while delivering low radiation doses and containing costs. The aim of this study was to assess, in terms of radiation dose and image quality, the possibility of using CBCT as an alternative to MSCT in patients with bionic ear implants. Materials and methods. One hundred patients (mean age 26 years, range 7–43) with Vibrant SoundBridge implants on the round window underwent follow-up: 85 with CBCT and 15 with MSCT. We measured the average tissueabsorbed doses during both MSCT and CBCT scans. Each scan was focused on the temporal bone with the smallest field of view and a low-dose protocol. In order to estimate image quality, we obtained data about slice thickness, high- and low-contrast resolution, uniformity and noise by using an AAPM CT performance phantom. Results. Although the CBCT images were qualitatively inferior to those of MSCT, they were sufficiently diagnostic to allow evaluation of the position of the implants. The effective dose of MSCT was almost three times higher than that of CBCT. Conclusions. Owing to low radiation dose and sufficient image quality, CBCT could be considered an adequate Riassunto Obiettivo. Gli impianti bionici all’orecchio rappresentano una soluzione contro la sordità. I pazienti trattati con questi tipi di ausili, spesso bambini, necessitano di un follow-up di esami sia funzionali che radiologici, in particolare mediante tomografia computerizzata multistrato (MSTC). La TC volumetrica dentale a fascio conico (CBCT) è una metodica utile allo studio dell’osso temporale, con il vantaggio di erogare basse dosi di radiazioni ionizzanti e di avere costi di gestione contenuti. Lo scopo dello studio è di valutare la CBCT in termini di dose radiante e di qualità d’immagine: il suo eventuale impiego in alternativa alla MSTC (con il protocollo attualmente utilizzato) comporterebbe infatti una sostanziale riduzione della dose ai pazienti con orecchio bionico. Materiali e metodi. Sono stati inclusi 100 pazienti ipoacusici (età media 26 anni, range 7–43), trattati mediante protesi Vibrant SoundBridge alla finestra rotonda, 15 studiati in follow-up mediante TC e 85 mediante CBCT, calcolando le dosi medie assorbite per tessuto sia durante esame MSTC che CBCT: ogni studio era focalizzato sull’osso temporale utilizzando il più piccolo campo di vista ed il protocollo a bassa dose radiante. Per la stima della qualità dell’immagine, sono stati ottenuti i valori di spessore di strato, risoluzione ad alto e basso contrasto e uniformità/rumore mediante l’uso di fantoccio AAPM CT performance. Risultati. Le immagini ottenute con la CBCT sono di qualità inferiore rispetto alla MSTC, ma sufficientemente diagnostiche, permettendo di valutare con sicurezza la Radiol med technique for postoperative imaging and follow-up of patients with bionic ear implants. Keywords Bionic ear implants · Cone-beam computed tomography · Radiation dose posizione delle estremità degli impianti acustici. La dose efficace della MSTC è risultata circa tre volte superiore rispetto a quella della CBCT. Conclusioni. La CBCT, grazie alla bassa dose impartita e alla sufficiente qualità delle immagini, può essere considerata tecnica radiologica adeguata per i controlli postoperatori ed il follow-up dei pazienti con orecchio bionico. Parole chiave Orecchio bionico, impianti acustici · Tomografia computerizzata a fascio conico · Dose radiante Introduction Introduzione Cone-beam computed tomography (CBCT) is an evolution of conventional computed tomography (CT) from which it has inherited some physical features, such as the use of ionising radiation, the ability to provide high-quality images and the possibility of imaging anatomical structures with multiplanar and three-dimensional (3D) reconstructions. It differs from its parent technique in its small size, low cost, limited field of view and lower radiation dose to the patient [1–4]. The first applications of CBCT were in the fields of dental and maxillofacial imaging, where it replaced in part orthopantomography and multislice CT (MSTC), which were limited by insufficient spatial information and high radiation doses, respectively. CBCT has played a primary role in dental implantology owing to its ability to provide an estimate of the thickness and depth of residual alveolar bone and visualise anatomical structures at risk during surgery, such as the mandibular canal, the mental foramen, the nasopalatine duct and the maxillary sinus. CBCT has also been employed to study the temporomandibular joints, bone neoplasms and single dental elements [5–7]. The use of CBCT is establishing itself in the otological field, where MSCT is the gold standard [8, 9]. The first clinical indications were middle-ear diseases causing conductive hearing impairment, in which it was used to identify the damaged conduction structure or to follow-up patients after ossiculoplasty [10]. CBCT was subsequently used for postoperative imaging and follow-up of patients with Vibrant SoundBridge (VSB) electromagnetic implants, both on the incus and on the round window [11, 12]. CBCT may offer a valuable alternative to MSCT in VSB hearing implants, as the patients – most of whom are children – require periodic follow-up, and CBCT delivers lower radiation doses. The purpose of this study was to evaluate a cohort of hearing-impaired patients with middle ear implants to determine how much dose could be saved by studying the La cone beam computed tomography (CBCT) o tomografia a fascio conico rappresenta una evoluzione della tomografia computerizzata convenzionale (TC), condividendone caratteristiche di natura fisica, come l’utilizzo di radiazioni ionizzanti, la capacità di ottenere immagini di elevata qualità, e la possibilità di studiare le varie strutture anatomiche con ricostruzioni multiplanari e tridimensionali; si differenzia dalla progenitrice TC per le dimensioni ridotte, i costi contenuti, il limitato campo di vista e la minore dose radiante somministrata al paziente [1–4]. Le prime applicazioni della CBCT sono state realizzate in campo odontoiatrico e maxillo-facciale, in parte sostituendo l’ortopantomografia e la TC multistrato (MSTC), per le quali l’insufficiente informazione spaziale per la prima, e la dose radiante per la seconda rappresentavano dei limiti. La CBCT ha rivestito un ruolo primario nell’implantologia dentale, per la capacità di fornire una stima dello spessore e della profondità dell’osso alveolare residuo e nell’evidenziare strutture anatomiche a rischio durante le procedure chirurgiche, come il canale mandibolare, il foro mentoniero, il dotto naso-palatino ed il seno mascellare; inoltre, la CBCT trova impiego anche nella valutazione delle articolazioni temporo-mandibolari, delle lesioni neoplastiche dell’osso, e dei singoli elementi dentari [5–7]. L’utilizzo della CBCT si sta imponendo in campo otologico, dove la MSTC rappresenta il gold standard [8, 9]. Le prime indicazioni cliniche hanno riguardato patologie dell’orecchio medio che determinavano ipoacusia di tipo trasmissivo, al fine di comprendere quale struttura di conduzione fosse lesa, o per il follow-up dei pazienti sottoposti ad intervento di ossiculoplastica [10]. Successivamente la CBCT è stata impiegata per l’imaging postoperatorio ed il follow-up dei pazienti impiantati con protesi elettromagnetiche Vibrant SoundBridge (VSB) sia all’incudine che alla finestra rotonda [11, 12]. La CBCT può rappresentare una valida alternativa allo studio MSTC in questo tipo di Radiol med petrous bones with CBCT rather than with MSTC, as well as to evaluate image quality obtained with the two techniques. Materials and methods Patients The study was conducted on 100 hearing-impaired patients (mean age 26 years, range 7–43) who, after initial clinical and aetiological assessment at our hospital’s Ear, Nose and Throat department, were treated with VSB implants on the round window. The time period considered spanned from May 2004 to November 2007. Postoperative follow-up included periodic functional tests to demonstrate restoration of hearing and radiological examinations to define the position of the implant and identify possible dislocations where clinically suspected. Patients underwent conventional radiology (anteroposterior and lateral skull radiography) and/or computed tomography (MSCT/CBCT). Equipment CBCT images were obtained with Maxiscan equipment (QR-DVT 9000, Verona, Italy), which generates a pulsed emission of a fixed 110-kVp conical beam of X-rays rotating around the patient’s head and is capable of reconstructing a cylindrical volume with a diameter and height of 15 cm. The 400-cm2-wide detector is made up of an image intensifier coupled with a CCD camera [1, 2, 13, 14]. Raw data are processed with primary reconstruction to obtain sections in the axial plane or other planes specified by the operator of a nominal thickness of 1 mm (in the protocol used). On the basis of the anteroposterior and lateral projection tomograms, CBCT calculates the milliampere per second (mAs) values for the axial scan (generally between 100 and 150 mAs). MSCT images of the petrous bones were obtained with a 120-kVp, 60-mAs, low-exposure protocol in the transverse plane with electronic reconstruction in the coronal plane (thickness 1 mm; spiral technique; pitch 0.4) using a Brilliance CT 6 MSCT device (Philips, Eindhoven, The Netherlands). Measurement of values of exposure to ionising radiation A comparison between the doses delivered to the patients and the associated radiological risk was performed with both devices. Measurements were obtained by placing 46 thermoluminescent dosimeters (TLDs) inside the head of a RANDO anthropomorphic phantom (Alderson Research impianti acustici con un risparmio di dose in pazienti, per lo più pediatrici, che devono sottoporsi a controlli periodici. Lo scopo del presente lavoro è quello di valutare una coorte di pazienti ipoacusici impiantati all’orecchio medio per definire il risparmio di dose ottenibile eseguendo lo studio delle rocche petrose mediante CBCT rispetto all’esame MSTC e di valutare la qualità delle immagini ottenute con le due diverse tecniche. Materiali e metodi Pazienti Sono stati inclusi 100 pazienti ipoacusici (età media 26 anni, range 7–43), inizialmente inquadrati dal punto di vista clinico ed eziologico presso l’Istituto di Otorinolaringoiatria della nostra struttura e successivamente trattati mediante posizionamento di protesi VSB alla finestra rotonda. L’intervallo di tempo considerato è stato tra il Maggio 2004 ed il novembre 2007. Il follow-up post-operatorio di questi pazienti includeva l’esecuzione periodica di test funzionali, in grado di provare la ripresa della capacità uditiva, nonché di esami radiologici, per la valutazione della posizione della protesi e la sua eventuale dislocazione, qualora clinicamente sospettata. Questi pazienti sono stati sottoposti ad indagini di radiologia convenzionale (Rx cranio in antero-posteriore e latero-laterale) e/o tecniche tomografiche computerizzate (MSTC/CBCT). Apparecchiature Le immagini CBCT sono state ottenute utilizzando l’apparecchiatura Maxiscan (QR-DVT 9000, Verona, Italia) che genera una emissione pulsata di un fascio conico di raggi X, con tensione di 110 kVp, non modificabile, che ruota attorno alla testa del paziente, ed è in grado di ricostruire un volume cilindrico con diametro e altezza di 15 cm. Il rivelatore, con estensione di 400 cm2, è costituito da un intensificatore di brillanza accoppiato ad una videocamera con schermo CCD [1, 2, 13, 14]. I dati grezzi ottenuti sono rielaborati mediante ricostruzione primaria, ottenendo sezioni assiali o su altri piani indicati dall’operatore, con spessore nominale di 1 mm (nel protocollo da noi utilizzato). La CBCT in base ai tomogrammi in proiezione antero-posteriore e latero-laterale calcola i valori dei mAs per la scansione assiale (generalmente tra 100 e 150 mAs). Le immagini MSCT mirate alla rocca petrosa sono state ottenute, mediante protocollo a bassa esposizione da 120 kVp e 60 mAs, sul piano trasversale con ricostruzione elettronica sul piano coronale (spessore 1 mm; tecnica spirale; pitch 0,4), mediante apparecchiatura MSCT Brilliance CT 6 (Philips, Eindowen, Olanda). Radiol med Laboratories, Stanford, CN, USA). The TLDs were wrapped in polyethylene and inserted into the holes in the phantom at the sites of interest (crystalline lens, thyroid, cerebrum, hypophysis and marrow of the skull and of C1, C2 and C7 vertebrae and nuchal skin) and at other significant sites to determine the dose distribution. The TLDs, GR200A lithium fluorides (Central Research Laboratory, Beijing, China), were read with a PCL3 reader (Fimel, Vélizy, France) while the annealing process occurred inside an ETT oven (Fimel, Vélizy, France). The dosimeters were calibrated in terms of air kerma for doses between 5 mGy and 30 mGy using an X-ray tube with 80 kVp and 2.8 mm Al filtration, through comparison with an ionisation chamber. Three TLDs were not irradiated in order to allow subtraction of the background signal. The phantom was examined at different times with CBCT and MSCT, following the manufacturer’s protocols for the specific examination (Table 1). The TLDs readings were used to derive the air kerma values. Tissue dose was obtained by multiplying the air kerma values by the ratio of air–muscle mass-energy absorption coefficients [15]. The calculation was performed according to the latest guidelines of the International Commission on Radiological Protection (ICRP) [16, 17]. The equivalent doses delivered to the different organs (Ht, measured in millisievert (mSv)] were calculated by multiplying the dose obtained with the TLDs by the volume fraction of the irradiated organ (Dt, measured in mGy) and by the weight coefficient of ionising radiation (WR): the weight coefficient of radiation is considered equal to 1. The contribution of a single organ/tissue in determining overall patient risk during exposure to radiation was obtained by multiplying the equivalent dose (Ht) by the tissue-weight coefficient (WT). The summation of the weighted products of the single organs gives the value of effective dose (E, measured in mSv), which is the dose Table 1 Standard exposure parameters of multislice computed tomography (MSCT) and cone-beam computed tomography (CBCT) for the protocols used for middle-ear evaluation Potential difference (kV) Exposure time (mAs) Indicated dose (mGy) MSCT CBCT 120 60 11.8 110 226.13 9.03 Tabella 1 Parametri espositivi standard di MSTC e CBCT per i protocolli valutativi dell’orecchio medio Differenza di potenziale (kV) Tempo di posa (mAs) Dose indicata (mGy) MSTC CBCT 120 60 11,8 110 226,13 9,03 Misurazione dei valori di esposizione alle radiazioni ionizzanti Il confronto fra le dosi impartite ai pazienti ed il rischio radiologico associato è stato effettuato con entrambe le apparecchiature. Le misure sono state effettuate posizionando 46 dosimetri termoluminescenti (TLDs) all’interno della testa di un fantoccio antropomorfo RANDO (Alderson Research Laboratories, Stanford, CN, USA). I TLDs sono stati avvolti nel polietilene e quindi inseriti negli appositi fori del fantoccio sia in siti di interesse (cristallino, tiroide, cervello, ipofisi, midollo osseo delle ossa della teca e delle vertebre cervicali C1, C2 e C7, cute nucale) che in altre posizioni significative per la determinazione della distribuzione di dose. I TLDs, fluoruri di litio del tipo GR200A (Central Research Laboratory, Pechino, Cina), sono stati letti con lettore PCL3 (Fimel, Vélizy, Francia) mentre il processo di annealing (temperaggio) è stato eseguito con un forno ETT (Fimel, Vélizy, Francia). I dosimetri sono tarati in termini di kerma in aria per dosi comprese tra 5 mGy e 30 mGy, utilizzando un tubo radiologico con 80 kVp e filtrazione di 2,8 mm Al, mediante confronto con camera a ionizzazione. Tre TLDs non sono stati irradiati in modo da poter sottrarre il segnale di fondo. Il fantoccio è stato esaminato in momenti diversi alla CBCT e alla MSCT seguendo i protocolli predisposti dalle rispettive ditte fornitrici per l’esame in oggetto (Tabella 1). Dalla lettura dei TLDs sono stati ricavati i valori di kerma in aria; moltiplicando questi ultimi per il rapporto dei coefficienti di assorbimento di energia massici ariamuscolo [15] si è ottenuta la dose in tessuto. La determinazione è stata effettuata secondo le più aggiornate linee guida dell’International Commission on Radiology Protection (ICRP) [16, 17]. Le dosi equivalenti impartite ai diversi organi (Ht, misurato in mSv) sono state ricavate moltiplicando la dose ottenuta mediante TLDs per la frazione del volume dell’organo irradiato (Dt, misurato in mGy) e per il coefficiente di peso della radiazione ionizzante (WR): il coefficiente di peso per la radiazione è considerato uguale a 1. Il contributo di un singolo organo/tessuto nel determinare il rischio globale del paziente durante l’esposizione radiante viene ottenuto moltiplicando la dose equivalente (Ht) per il coefficiente di peso del tessuto (WT). La sommatoria dei prodotti pesati per i singoli organi fornisce il valore di dose efficace (E, misurato in mSv) che è la grandezza dosimetrica legata al rischio stocastico, somatico e genetico (E=兺Ht×Wt). Qualità dell’immagine Il confronto della qualità dell’immagine tra le due TC è stato effettuato utilizzando un fantoccio di forma cilindrica con Radiol med magnitude related to the stochastic, somatic and genetic risk (E=兺Ht×Wt). Image quality Image quality of the two CT techniques was compared using a dedicated cylindrical phantom measuring 21.6 cm in diameter (AAPM CT Performance Phantom 76-411, Victoreen Instruments, Mödling, Austria) equipped with inserts allowing image quality to be evaluated in terms of slice thickness, high- and low-contrast resolution and ratio between signal uniformity and noise. Axial images of the phantom inserts were acquired using CBCT and MSTC with the protocols described for dose measurements. The image obtained with CBCT measures 15 cm in diameter and therefore does not cover the whole phantom section. The slice thickness of the reconstructed image was determined by measuring its width at half the height of the profile of the lamellae fitted into the appropriate insert. To avoid sampling errors, average size was determined in different image positions. To allow determination of high-contrast spatial resolution, the plexiglas phantom contains eight sets of five round, air-filled holes between 2.5 mm and 0.75 mm in diameter and spaced in such a way that the distance between their centres is twice their diameters. The visibility of details was evaluated by selecting grey-scale with a minimal window and choosing the level at which the details were best identified. Details were considered visible if they were distinct. Contrast pins with a diameter of 1 in. (25.4 mm) were immersed in water inside the phantom. The pins have the following densities: polycarbonate = 1.20 g/cm3, acrylic = 1.19 g/cm3, nylon = 1.10 g/cm3, polystyrene = 1.05 g/cm3 and polyethylene = 0.95 g/cm3. These inserts can provide some indication of low-contrast resolution. When measuring uniformity, the value considered was the maximum ratio between the average pixel values in the regions of interest (ROI) placed at the centre and in four peripheral locations of an image of the homogeneous insert of the phantom. Finally, we assessed whether it was possible to visualise the various anatomical components of the tympanic cavity and specifically of the VSB implants to determine the position of the distal end at the level of the round window or of the long apophysis of the incus. Results All 100 patients underwent MSCT within 6 h of surgery, according to the protocol. During follow-up, 2/100 patients underwent skull radiography in orthogonal projections, 15/100 were studied with MSCT of the petrous bones and diametro di 21,6 cm dedicato (AAPM CT Performance Phantom 76-411, Victoreen Instruments, Mödling, Austria), dotato di inserti per valutare la qualità dell’immagine in termini di spessore di strato, di alta e bassa risoluzione di contrasto e rapporto uniformità/rumore del segnale. Mediante CBCT e MSTC sono state ricavate immagini assiali degli inserti del fantoccio, utilizzando i protocolli precedentemente descritti per la misurazione delle dosi. Per le scansioni con la CBCT l’immagine ottenuta ha diametro di 15 cm, quindi non comprende l’intera sezione del fantoccio. Lo spessore di strato dell’immagine ricostruita è stato determinato misurando la larghezza a metà altezza del profilo delle lamelle inserite nell’apposito inserto; per evitare l’errore dovuto al campionamento è stata determinata la media delle dimensioni in diverse posizioni dell’immagine. Per la determinazione della risoluzione spaziale ad alto contrasto, nel fantoccio in plexiglas sono presenti otto serie di cinque fori pieni d’aria a sezione circolare con diametro da 2,5 mm a 0,75 mm distanziati fra loro in modo tale che la distanza fra i centri sia doppia del diametro del foro. Per valutare la visibilità dei particolari è stata selezionata la scala di grigi con finestra ridotta al minimo scegliendo il livello che meglio permette di distinguerli; questi vengono considerati visibili se risultano tutti distinti. Nel fantoccio sono inseriti dei cilindri con diametro di un pollice (25,4 mm) immersi in acqua con le seguenti densità: policarbonato=1,20 g/cm3, acrilico=1,19 g/cm3, nylon=1,10 g/cm3, polistirene=1,05 g/cm3, polietilene=0,95 g/cm3. Questi inserti possono fornire una indicazione della risoluzione a basso contrasto. Per la misura dell’uniformità viene considerato il massimo rapporto fra i valori medi dei pixel nelle regioni di interesse (ROI) posizionate al centro ed in quattro posizioni alla periferia di una immagine acquisita in corrispondenza dell’inserto omogeneo del fantoccio. È stata infine valutata la possibilità di evidenziare le diverse componenti anatomiche presenti all’interno del cavo timpanico, e nel caso specifico delle protesi VSB, in particolare per definire con certezza la posizione dell’estremo distale a livello della finestra rotonda o all’apofisi lunga dell’incudine. Risultati Tutti i 100 pazienti sono stati sottoposti a esame MSTC entro le 6 ore dall’intervento, come da protocollo postchirurgico. Successivamente, in corso di follow-up, i pazienti che hanno eseguito Rx cranio nelle proiezioni ortogonali sono stati 2/100; i pazienti studiati mediante MSTC delle rocche petrose sono stati 15/100, e 85/100 mediante CBCT. L’analisi delle immagini di radiologia convenzionale non permetteva di fornire una precisa stima della Radiol med a b Fig. 1a,b Patient with a Vibrant SoundBridge device implanted on the round window. Axial multislice computed tomography (a) compared with the cone-beam computed tomography scan (b) showing the correct position of the floating mass. Fig. 1a,b Paziente portatore di impianto Vibrant Soundbridge alla finestra rotonda. Sezione assiale MSCT (a) e corrispondente immagine CBCT (b) che mostra il corretto posizionamento della massa vibrante. 85/100 were studied with CBCT. Conventional radiography failed to allow accurate definition of the position of the implant, whereas this proved possible with both crosssectional imaging techniques. Although the images obtained with CBCT were of lower quality, they were sufficiently diagnostic to permit confident reporting by the radiologist of the position of implant extremities (Fig. 1). The weighted equivalent doses for the single tissues and the effective dose obtained using the RANDO phantom after CBCT and MSCT are listed in Table 2. With regard to the crystalline lens, the measured dose was 4.5 mGy for CBCT and 15.5 mGy for MSCT, values that are substantially lower than those likely to cause deterministic effects, even after repeated exams. An overall analysis of the single results shows that the effective dose delivered to the individual patient during CBCT of the ear was equal to 39.2% of the dose imparted during MSCT. In particular, the effective MSCT dose was 0.28 mSv, whereas that of CBCT was 0.11 mSv. Results of the assessment of image quality are summarised in Table 3. In particular, slice thickness was 1 posizione della protesi, al contrario fornita da entrambe le tecniche tomografiche. Le immagini ottenute con la CBCT sono di qualità inferiore ma sufficientemente diagnostiche per il medico radiologo che deve refertare l’esame, permettendo di valutare con sicurezza la posizione delle estremità degli impianti acustici (Fig. 1). Le dosi equivalenti ponderate per i singoli tessuti e la dose efficace ottenute mediante il fantoccio RANDO dopo scansione CBCT e MSTC sono elencate in Tabella 2. Per quanto riguarda il cristallino la dose è risultata pari a 4,5 mGy per la CBCT ed a 15,5 mGy per la MSTC, valori decisamente inferiori a quelli che possono provocare effetti deterministici anche per esami ripetuti. Da una analisi globale dei singoli risultati si evince come la dose efficace somministrata ad un paziente durante esame CBCT dell’orecchio sia il 39,2% rispetto a quella di un’indagine MSTC: nello specifico la dose efficace della MSTC è stata di 0,28 mSv, mentre quella della CBCT è stata di 0,11 mSv. Per quanto riguarda gli studi mirati a valutare la qualità dell’immagine i dati sono riassunti in Tabella 3. In partico- Table 2 Tissue and organ equivalent and effective doses for multislice computed tomography (MSCT) and cone-beam computed tomography (CBCT) MSCT CBCT Organ/tissue Absorbed dose (mGy) % vol. Wt Weighted equivalent dose (mSv) Absorbed dose (mGy) % vol. Wt Weighted equivalent dose (mSv) Bone marrow Oesophagus Thyroid Bone surface Brain Salivary glands Skin Other 7.01 16.5 10 100 16.5 20 100 5 5 0.12 0.05 0.05 0.01 0.01 0.01 0.01 0.1 0.14 0.00 0.03 0.01 0.02 0.04 0.01 0.03 2.27 0.32 2.23 2.97 2.52 3.16 2.42 16.5 10 100 16.5 20 100 5 5 0.12 0.05 0.05 0.01 0.01 0.01 0.01 0.1 0.04 0.00 0.02 0.00 0.01 0.03 0.00 0.01 0.28 Total effective dose (mSv) 0.60 7.50 7.92 4.16 11.41 6.78 Total effective dose (mSv) MSCT, multislice computed tomography; CBCT, cone-beam computed tomography, Wt , tissue-weight coefficient 0.11 Radiol med Tabella 2 Dosi equivalenti agli organi e tessuti e dose efficace per le due tecniche TC MSTC Midollo osseo Esofago Tiroide Sup. ossea Encefalo Ghiandole salivari Pelle Rimanenti CBCT Dose assorbita (mGy) % vol Wt Dose equivalente pesata (mSv) Dose assorbita (mGy) % vol Wt Dose equivalente pesata (mSv) 7,01 16,5 10 100 16,5 20 100 5 5 0,12 0,05 0,05 0,01 0,01 0,01 0,01 0,1 0,14 0,00 0,03 0,01 0,02 0,02 0,01 0,03 2,27 0,32 2,23 2,97 2,97 3,16 2,42 16,5 10 100 16,5 20 20 5 5 0,12 0,05 0,05 0,01 0,01 0,01 0,01 0,1 0,04 0,00 0,02 0,00 0,01 0,01 0,00 0,01 0,28 Dose efficace totale (mSv) 0,60 7,50 7,92 4,16 11,41 6,78 Dose efficace totale (mSv) 0,11 MSTC, tomografia computerizzata multistrato; CBCT, tomografia computerizzata a fascio conico; Wt , coefficiente di peso del tessuto Table 3 Results of the image quality evaluation by means of the AAPM CT performance phantom Slice thickness (mm) High-contrast resolution (pl/cm) No. of visible low-contrast inserts (out of 5) Uniformity/noise ratio MSCT CBCT 1 4 5 1 3 2 1% 27% MSCT, multislice computed tomography; CBCT, cone-beam computed tomography; pl/cm, pair of lines per centimetry Tabella 3 Risultati della valutazione della qualità delle immagini mediante fantoccio AAPM CT performance Spessore dello strato (mm) Risoluzione alto contrasto (pl/cm) N. inserti a basso contrasto visibili (su 5 presenti) Rapporto uniformità/rumore MSTC CBCT 1 4 5 1 3 2 1% 27% MSCT, tomografia computerizzata multistrato; CBCT, tomografia computerizzata a fascio conico; pl/cm, paia di linee per centimetro mm on both CBCT and MSCT (Fig. 2). In the evaluation of high-contrast spatial resolution, detail was well-defined down to 1 mm on both CBCT and MSCT (Fig. 3). With regard to low-contrast spatial resolution, CBCT proved to be more limited than MSCT, with CBCT accurately visualising two details out of five and MSCT depicting all five (Fig. 4). Uniformity was worse on CBCT (27% as against 1% on MSCT) (Fig. 5). In addition, CBCT was affected by artefacts appearing as circular streaks, which were absent from the MSCT images. lare, lo studio dello spessore dello strato è risultato essere di 1 mm sia con la CBCT che con la MSTC (Fig. 2). Nella valutazione della risoluzione spaziale ad alto contrasto, i particolari risultano tutti ben distinguibili fino a dimensioni di 1 mm sia con la CBCT che con la MSTC (Fig. 3). Nelle misurazioni per la risoluzione spaziale a basso contrasto la CBCT ha dimostrato dei limiti rispetto alla MSTC: con la CBCT sono visibili con certezza due particolari su cinque, mentre con la MSTC sono visibili tutti i cinque particolari (Fig. 4). L’uniformità è peggiore per la CBCT (27% contro 1% della MSTC) (Fig. 5), ed inoltre all’esame visivo alla CBCT risultano presenti degli artefatti che si manifestano come striature ad andamento circolare, non presenti nell’immagine MSTC. Discussione Con questo studio si è cercato di determinare quale fosse la metodica in grado di fornire il miglior rapporto qualità diagnostica/dose impartita al paziente nella valutazione del corretto posizionamento della protesi acustica. Con queste premesse, abbiamo considerato l’eventualità di sottoporre i pazienti operati ad un follow-up radiologico che sfruttasse tecniche alternative alla MSTC: poiché metodiche non ionizzanti non erano usufruibili, la radiografia standard del cranio e la CBCT rappresentavano le uniche opzioni. Tutti i 100 pazienti della nostra casistica sono stati valutati nell’immediato post-operatorio come da protocollo (entro le 6 ore) mediante esame MSTC al fine di escludere focolai emorragici in sede di intervento o un eventuale mal posizionamento della protesi. Il follow-up standard includeva l’esecuzione di una indagine MSTC a 6 settimane, ossia all’attivazione dell’impianto elettromagnetico, ed a 1 Radiol med a b Fig. 2a,b Slice thickness for multislice computed tomography (a) and cone-beam computed tomography (b) defines the effective tomographic section and depends both on the collimation of the outgoing beam and on a subsequent collimation at the detectorsystem level; b shows a lower contrast with the same slice thickness. Fig. 2a,b Lo spessore di strato per la MSTC (a) e per la CBCT (b) definisce l’effettiva sezione tomografica, ed è determinato dalla collimazione del fascio in uscita dal tubo radiogeno e da una post collimazione al livello dei rivelatori: in B è evidente come a parità di spessore dello strato sia minore il contrasto. a b Fig. 3a,b High-contrast resolution, that is, the system’s ability to distinguish small-sized objects against a high-contrast background, was only slightly higher for multislice computed tomography (a) than cone-beam computed tomography (b). Fig. 3a,b La risoluzione ad alto contrasto, intesa come la capacità del sistema di distinguere oggetti di piccole dimensioni in un contesto di alto contrasto, è risultata leggermente superiore per la MSTC (a), rispetto alla CBCT (b). a b Fig. 4a,b Low-contrast resolution on MSCT (a) and CBCT (b) refers to the system’s ability to distinguish small-sized objects against a low-contrast background. MSCT clearly offers higher discrimination ability than CBCT. Fig. 4a,b La risoluzione a basso contrasto per la MSTC (a) e per la CBCT (b) è intesa come la capacità del sistema di distinguere oggetti di piccole dimensioni in un contesto di basso contrasto. È evidente la migliore capacità discriminativa della MSTC rispetto alla CBCT. Discussion This study sought to identify the technique providing the best ratio between diagnostic quality and patient dose in the evaluation of the correct position of VSB ear implants. To this end, patients were offered a radiological follow-up anno di distanza; eventuali controlli successivi sarebbero stati effettuati solo su fondate alterazioni cliniche. Quindici/100 pazienti hanno seguito l’iter tradizionale, e quindi sono stati studiati mediante MSTC. Il follow-up da noi proposto implicava l’esecuzione di un Rx cranio e/o di una CBCT rispetto alla MSTC: i primi Radiol med a b Fig. 5a,b Uniformity is defined as the maximum ratio between the average values of pixels in different regions of interest, with one placed at the centre of the image and four in peripheral locations. Uniformity with multislice computed tomography (a) is clearly higher than with cone-beam computed tomography (b). Fig. 5a,b L’uniformità è definita come il massimo rapporto fra i valori medi dei pixel di regioni di interesse poste una al centro e quattro in periferia, con la MSTC (a) e con CBCT (b). La MSCT è visibilmente più uniforme della CBCT. consisting of standard skull radiography and CBCT as alternatives to MSCT, given that no nonionising modalities were available. All 100 patients in our series underwent postoperative evaluation with MSCT within 6 h of surgery to exclude bleeding at the operation site or incorrect placement of the implant. The standard follow-up protocol includes MSCT at 6 weeks – that is, on activating the electromagnetic implant – and again after 1 year. Any further assessment is done only in the case of demonstrated clinical changes. Fifteen out of 100 patients were followed up with the standard protocol (MSCT). The experimental follow-up protocol involved performing skull radiography and/or CBCT instead of MSCT. The first 2/100 patients underwent skull radiography in orthogonal projections, but this strategy was abandoned due to the lack of adequate landmarks. The following 85/100 patients were studied with CBCT, and with them, we attempted to validate the method and familiarise ourselves with the images. Finally, CBCT was considered sufficient so that MSCT could be reserved for cases suspected to have more severe complications than implant dislocation (no case in the period in question). Analysis of dosimetric data, in agreement with the literature [18–22], showed that the use of CBCT permits considerable dose saving compared with MSCT. This is made possible by beam geometry, type of emission (pulsed rather than continuous over the whole tube-rotation angle) and more efficient detection. In addition, although CBCT provides lower-quality images, these are nonetheless adequate for evaluating the position of the hearing implant. In contrast to MSCT, CBCT scans and reconstructions are not affected by artefacts related to the metallic components of the vibrating mass of the implant, which often limit visualisation of the structures being investigated. CBCT confirms its good high-contrast resolution, but unlike MSCT, it is inadequate at low contrast. These two charac- 2/100 pazienti hanno espletato indagine Rx del cranio nelle proiezioni ortogonali, dalla quale non si riuscivano ad estrapolare sufficienti reperi, e quindi è stata abbandonata e non riproposta nei successivi pazienti. Ottantacinque/100 pazienti sono stati sottoposti a esame CBCT e con essi si è cercato di validare la metodica e prendere confidenza con le immagini ottenute. Si è infine ritenuto sufficiente l’esecuzione di questa sola metodica, riservando la MSTC a quelle circostanze in cui il sospetto fosse di complicazioni più gravi della semplice dislocazione della protesi (nessun caso nella nostra casistica nel periodo in esame). L’analisi dei dati dosimetrici, in accordo con la letteratura [18–22], dimostra come l’utilizzo della CBCT comporti una considerevole riduzione di dose alla MSTC, determinata in particolare dalla geometria del fascio, dal tipo di emissione (pulsata, rispetto ad un’emissione continua su tutto l’angolo di rotazione del tubo) e da una maggiore efficienza di rilevazione. Oltre a ciò, la CBCT evidenzia una riduzione della qualità dell’immagine ottenuta, che comunque rimane sufficiente per la risoluzione del problema diagnostico della valutazione del posizionamento protesico. A differenza della MSTC, la CBCT non presenta nelle scansioni e nelle successive ricostruzioni artefatti derivanti dalle componenti metalliche della massa vibrante dell’impianto acustico, procuranti spesso limitazioni nella visualizzazione delle strutture d’interesse. La CBCT conferma di avere una buona risoluzione ad alto contrasto ma a differenza della MSTC risulta deficitaria nel basso contrasto: queste due caratteristiche conferiscono alla CBCT la capacità di visualizzare con buona risoluzione di contrasto le strutture ossee, mentre con difficoltà i tessuti molli. L’orecchio medio è una struttura anatomica, inserita in un contesto osseo, la rocca petrosa dell’osso temporale, e riempita di aria attraverso la tuba di Eustachio; al suo interno la trasmissione del suono è garantita dalla catena ossiculare, formata da singoli ossicini. La CBCT riesce a evidenziare correttamente le diverse componenti anato- Radiol med teristics mean that CBCT visualises bone structures with good contrast resolution, whereas it is somewhat limited in depicting soft tissues. The middle ear is contained within a bony context – the petrous portion of the temporal bone – filled with air through the Eustachian tube. Sound conduction is ensured by the chain of ossicles. CBCT correctly depicts the various anatomical components within the tympanic cavity and, in the case of VSB implants, allows accurate definition of the position of the distal end at the level of the round window or of the long apophysis of the incus. Because CBCT has been reported to impart smaller radiation doses compared with MSCT, we attempted to determine the difference in dosimetric performance between the devices available at our department. Data analysis showed that CBCT delivers a patient dose equal to 39% of that imparted by MSCT with a low-dose protocol. According to the directive of the Dimond III European research group, when choosing between two techniques capable of solving a diagnostic problem, one needs to take into account both its usefulness in depicting the object being studied and the patient dose [23]. In the case of VSB devices, it is important to consider that the majority of patients are children, and they have to undergo a intensive postoperative follow-up protocol involving repeated radiological examinations. Such patients are those most likely to be affected by the cumulative effects of ionising radiation. CBCT meets the European criteria for diagnostic capability, and its place in otology – where MSCT is still the gold standard – should be reassessed. The limitations of this study, in which the comparison was based on the standard protocols in use at our hospital and on the best cost–benefit ratio for each of the CT scanners available at the time of assessment, concern the unexpressed potential for reducing the MSCT dose by refining the protocol. However, there remains an objective difference between the doses delivered to individual patients, even after attempting to stress the protocols at a dosimetric level, which is due to the intrinsic characteristics of the two CT applications. We therefore believe that our study may encourage a more widespread use of CBCT in other fields as well, given that while yielding the same diagnostic information, CBCT allows lower patient doses and lower costs per examination. miche presenti all’interno del cavo timpanico e, nel caso specifico delle protesi VSB, a definire con certezza la posizione dell’estremo distale a livello della finestra rotonda o all’apofisi lunga dell’incudine. Dalla letteratura si conferma come la CBCT somministri una quantità di radiazioni ionizzanti minore rispetto alla MSTC; si è quindi voluto ottenere un riscontro oggettivo della differenza di prestazioni dosimetriche fra le apparecchiature a disposizione presso il nostro Dipartimento. A conclusione dei rilievi e dell’elaborazione dei dati si rileva come la CBCT eroghi una dose radiante pari al 39% di quella della MSTC con protocollo a bassa esposizione. Secondo la raccomandazione del gruppo di ricerca europeo Dimond III nella scelta fra due tecniche bisogna considerare sia l’utilità della stessa nel mettere in evidenza l’oggetto di studio, sia la dose radiante, se il quesito diagnostico viene soddisfatto [23]. Nel caso delle protesi VSB si deve tenere conto che spesso i pazienti sono in età pediatrica e che dal momento dell’intervento entrano a far parte di un protocollo di follow-up radiologico intenso, composto da controlli ripetuti; sono proprio questi pazienti che più potrebbero risentire degli effetti cumulativi delle radiazioni ionizzanti. La CBCT rispetta i criteri europei di capacità diagnostica e andrebbe rivalutata la sua posizione nella diagnostica in campo otologico, dove la MSTC rappresenta ancora il gold standard. I limiti di questo lavoro, nel quale il confronto è stato compiuto secondo i protocolli di comune uso nella pratica clinica nella nostra struttura, e secondo il miglior rapporto costo-beneficio per ognuna delle macchine a tecnologia TC possedute al momento della valutazione dei pazienti, riguardano la potenzialità non espressa di diminuire la dose in MSCT, affinando il protocollo: resta comunque oggettivo il divario tra le dose impartite ai singoli pazienti, anche cercando di stressare dosimetricamente i protocolli, proprio per le caratteristiche intrinseche delle due diverse applicazioni di tecnologia TC, e per questo noi riteniamo che questo lavoro possa comunque essere di stimolo per cercare di diffondere l’uso della CBCT anche in altre applicazioni, in modo da ridurre la dose al paziente, poiché in caso di parità di informazione diagnostica si ha una minor dose al paziente e un minor costo per esame. Conclusioni Conclusions During CBCT of the ear, the radiation dose delivered to the patient is lower than that of MSCT performed with a lowdose protocol. In comparison with MSCT, CBCT provides an image quality that is lower in terms of low-contrast resolution and uniformity and slightly lower in terms of high- Durante un esame CBCT dell’orecchio la dose radiante somministrata al paziente è inferiore a quella erogata dalla TC con protocollo a bassa esposizione. La qualità dell’immagine ottenuta con CBCT, rispetto alla MSTC, è minore in termini di risoluzione a basso contrasto e uniformità, è leggermente inferiore per risoluzione ad alto contrasto, mentre è equivalente per quanto riguarda lo Radiol med contrast resolution, whereas slice thickness is equal. The images are sufficiently diagnostic to identify the correct placement of VSB, both on the round window and on the incus, with the advantage of lower radiation exposure for the patient. CBCT may candidate itself to be a first-line technique in the follow-up of patients with middle-ear implants. spessore dello strato. Le immagini risultano sufficientemente diagnostiche per identificare il corretto posizionamento del VSB sia alla finestra rotonda che all’incudine, con il vantaggio di esporre il paziente a minori rischi da esposizione a radiazioni ionizzanti: la CBCT può candidarsi a metodica di primo livello nel follow-up dei pazienti con protesi all’orecchio medio. Conflict of interest statement The authors declare that they have no conflict of interest to the publication of this article. References/Bibliografia 1. Mozzo P, Procacci C, Tacconi A et al (1998) A new volumetric CT machine for dental imaging based on the conebeam technique: preliminary results. Eur Radiol 8:1558–1564 2. Robinson S, Suomalainen A, Kortesniemi M (2005) µ-CT. Eur J Radiol 56:185–191 3. Ludlow JB, Davies-Ludlow LE, Brooks SL et al (2006) Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT. Dentomaxillofac Radiol 35:219–226 4. Mah JK, Danforth RA, Bumann A et al (2003) Radiation absorbed in maxillofacial imaging with a new dental computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 96:508–513 5. Ludlow JB, Laster WS, See M et al (2007) Accuracy of measurements of mandibular anatomy in cone beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:534–542 6. Tsiklakis K, Donta C, Gavala S et al (2005) Dose reduction in maxillofacial imaging using low dose Cone Beam CT. Eur J Radiol 56:413–417 7. Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus Ds panoramic unit. Dentomaxillofac Radiol 32:229–234 8. Peltonen LI, Aarnisalo AA, Kortesniemi MK et al (2007) Limited cone-beam computed tomography imaging of the middle ear: a comparison with multislice helical computed tomography. Acta Radiol 48:207–212 9. Dalchow CV, Weber AL, Yanagihara N et al (2006) Digital volume tomography: radiologic examinations of the temporal bone. AJR Am J Roentgenol 186:416–423 10. Dalchow CV, Weber AL, Bien S et al (2006) Value of digital volume tomography in patients with conductive hearing loss. Eur Arch Otorhinolaryngol 263:92–99 11. Colletti V, Soli SD, Carner M et al (2006) Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. Int J Audiol 45:600–608 12. Cerini R, Faccioli N, Barillari M et al (2008) Bionic ear imaging. Radiol Med 113:265–277 13. Araki K, Maki K, Seki K et al (2004) Characteristics of a newly developed dentomaxillofacial X-ray cone beam CT scanner (CB MercuRay): system configuration and physical properties. Dentomaxillofac Radiol 33:51–59 14. Lascala CA, Panella J, Marques MM (2004) Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCTNewTom). Dentomaxillofac Radiol 33:291–294 15. Johns HE, Cunningham JR (1983) The physics of radiology, 4th edn. Charles C Thomas Publisher, Springfield Illinois USA 16. International Commission on Radiological Protection (1991) 1990 Recommendations of the International Commission on Radiological Protection, ICRP Publication 60. Ann ICRP 21:1–201 17. International Commission on Radiological Protection (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37:1–332 18. Cohnen M, Kemper J, Mobes O et al (2002) Radiation dose in dental radiology. Eur Radiol 12:634–637 19. Gonzalez L, Vaño E, Fernandez R (2001) Reference doses in dental radiodiagnostic facilities. Br J Radiol 74:153–156 20. Diederichs CG, Engelke WG, Richter B et al (1996) Must radiation dose for CT of the maxilla and mandible be higher than that for conventional panoramic radiography? AJNR Am J Neuroradiol 17:1758–1760 21. Lecomber AR, Yoneyama Y, Lovelock DJ et al (2001) Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and computed tomography. Dentomaxillofac Radiol 30:255–259 22. Schulze D, Heiland M, Thurmann H et al (2004) Radiation exposure during midfacial imaging using 4- and 16-slice computed tomography, cone beam computed tomography systems and conventional radiography. Dentomaxillofac Radiol 3:83–86 23. Busch HP, Faulkner K (2005) Image quality and dose management in digital radiography: a new paradigm for optimization. Radiat Prot Dosimetry 117:143–147